    ISO 8301 – Thermal conductivity testing

    Thermal insulation - steady-state thermal resistance - heat flow meter apparatus

    Testing · NAC-037-TL

    In short

    ISO 8301 - Thermal insulation - steady-state thermal resistance - heat flow meter apparatus. Amstelfy's laboratory determines thermal resistance and conductivity (heat flow meter) of thermal insulation products (MW, EPS, XPS, PIR, PUR, wood fibre) and building materials according to ISO 8301 under its ISO/IEC 17025 accreditation (NAC-037-TL). The method is typically part of initial type testing, ETA test programmes or factory production control verification; results are reported in English for manufacturers across Europe.

    What ISO 8301 determines

    ISO 8301 describes the heat flow meter apparatus for measuring steady-state thermal resistance. A flat specimen is held between a hot plate and a cold plate at fixed temperatures, and one or two heat flux transducers record the heat flowing through it once the temperatures and the signal have stopped drifting. The apparatus is a relative instrument: the transducer output is converted into heat flux by calibration against reference specimens whose resistance is known from guarded hot plate measurements, so the quality of the calibration sets the quality of the result. The method yields the thermal resistance of the specimen and, for homogeneous products, its thermal conductivity at the mean test temperature. It is the instrument most laboratories use for routine conductivity work on insulation because it reaches equilibrium faster than an absolute guarded hot plate.

    Specimens, conditioning and procedure

    Specimens are flat slabs cut to the plate size of the apparatus, with parallel faces and a thickness inside the range for which the instrument was calibrated. They are conditioned to constant mass, usually at standard laboratory climate, and their thickness, mass and density are recorded. The operator sets the hot and cold plate temperatures to give the required mean temperature and a temperature difference within the calibrated range, then waits for steady state, confirmed by successive readings that agree within the drift limit. Options the client decides are the mean temperature (10 °C for European declarations, other values for export markets), single- or two-specimen configuration and whether soft products are tested at a defined thickness under the plate pressure.

    What the report states

    The report gives thermal resistance R in m²·K/W and, for homogeneous specimens, thermal conductivity λ in W/(m·K) at the stated mean temperature, with specimen thickness, density, conditioning state and the calibration references used. Several specimens from one product give a mean value; declared values for CE marking are derived from such data by the statistical procedure in the product standard, not by the test method itself.

    Where the result is required

    The heat flow meter provisions of EN 12664 and EN 12667 are built on ISO 8301, so results feed the declared thermal resistance and conductivity (λD, RD) of mineral wool, EPS, XPS, PIR, PUR and wood fibre products under EN 13162, EN 13163, EN 13164 and their sibling standards. The method also serves factory production control checks, product development, conductivity of sandwich panel cores for EN 14509 calculations, and specifications in markets that cite ISO rather than EN standards.

    What we assess

    • thermal resistance and conductivity (heat flow meter)

    Products covered

    • thermal insulation products (MW, EPS, XPS, PIR, PUR, wood fibre)
    • building materials

    Accreditation status

    ActivityStandardAccreditationValidityOn scope
    TestingISO/IEC 17025:2017NAC-037-TLvalid until 2 December 2028Yes
    Product certificationISO/IEC 17065:2012NAC-026-PPSvalid until 31 January 2030No
    Validation & verificationISO/IEC 17029:2019NAC-006-VVvalid until 31 January 2030No

    Route to market

    This test method is used inside initial type testing, ETA assessment programmes and factory production control verification. Amstelfy reports it as an accredited test under NAC-037-TL, either stand-alone or as part of a complete test programme.

    How it works

    1. 1

      Send the standard number, product description and intended use

    2. 2

      We confirm the applicable test series, sample quantities and timeline

    3. 3

      Samples are delivered to Veilmeesterweg 20, 1187 ZW Amstelveen (Mon-Fri 09:00-17:00)

    4. 4

      You receive the report in English, written for the TAB, Notified Body, evaluation service or programme operator.

    Frequently asked questions

    Yes - it is listed on the ISO/IEC 17025 testing scope NAC-037-TL.

    The transducer gives a voltage, not a heat flux. Its sensitivity depends on temperature and on contact conditions, so it is calibrated with reference specimens of known thermal resistance, traceable to guarded hot plate measurements. The specimen tested should fall within the resistance and thickness range covered by that calibration; otherwise the uncertainty rises.

    European product standards declare λ at a mean temperature of 10 °C, so that is the default. Other mean temperatures are measured on request, for example to build a conductivity–temperature curve or to meet specifications outside Europe; the temperature difference across the specimen is kept within the calibrated range.

    One specimen gives a single value; a declared value requires a set of measurements spread over production. For an initial measurement, a few boards large enough to cut specimens matching the apparatus plate size, at the thickness to be declared, are enough. Thin or very light products may need stacking or the separate thick-product procedure.